FEATURES
1. The instrument has two independent detecting places. It can detect different items or different samples of the same item contemporary.
2. It is based on photoelectric principle and can detect items of PT, APTT, TT, FBG and active measurements of many cruor factors.
3. It has a device and reagent-adding can be automatically detected by which. You can use ordinary sucker to add regents. For your convenience, the mixing device is vibratile so you do not need to beat it up.
Notice: We do not provide reagents so you should buy them yourselves, and the reagent must be cruor reagent and be enrolled in your country.
4. It has an inner printer so you can conveniently print the detecting report.
5. There are advanced design of optic system and data processing software to eliminate the disturbances of high-fat-blood, icterus, hemolysis and chyle.
6. It has many functions such as: detection, set, inquiry and showing & saving test results.
7. It has heating-timing function.
8. It can save at most 200 test results per day.
9. It can delete test results of the last day when it starts up for the first time this day.
SPECIFICATIONS
1. The driftance of transmission should be less than 1%;
2. When the transmission ratio is about 30%, measuring accuracy error should be less than 1.5%;
3. When the ratio of transmission is about 30%, error of measuring accuracy between channels should be less than 3%;
4. When the ratio of transmission is about 30%, error of measuring repeat between channels should be less than 0.5%;
5. When the ratio of transmission ranges from 10% to 80%, linear error of absorbance should be less than 2%;
6. The temperature of detecting places and warm-up places is 37℃. Error of temperature accuracy should be less then 0.2℃ and fluctuation of temperature should be less then 0.3℃. The temperature of refrigerated places ranges from 2℃ to 8℃;
7. Repeat error of PT detection should be less than 8% using the same sample in the same channel;
Work Conditions
1. Environmental temperature: 10℃-30℃;
2. Relative humidity: less than 70%;
3. Power supply: 220V±10%, 50Hz±2%;
4. Input power: 150 VA±10%;
5. The instrument should keep away from strongly electromagnetism, and avoid dust, shake and caustic gas;
6. Power supply should be earthed perfectly;
7. The instrument should be warm 20 minutes before operation